Passive Mechanical Skin Stretch for Multiple Degree-of-Freedom Proprioception in a Hand Prosthesis
Year: 2014
Type: conference-paper
Abstract: In this paper, we present a passive linear skin stretch device that can provide proprioceptive feedback for multiple degrees of freedom (DOF) in a prosthetic hand. In a 1-DOF virtual targeting task, subjects performed as well with our device as with a vibrotactile array, and significantly better ( $$p<0.05$$ ) than having no feedback at all. In a 3-DOF grip recognition task, subjects were able to classify six different grips with 88.0 % accuracy. Training took 6 min and the average time to classification was 5.2 s. Subjects were also able to match a set of target grip apertures with 11.1 % error on average.
Institution: University of Illinois Urbana-Champaign
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $5,392
OpenAlex ID: W1006714145