Diseño mejorado de bobinas de gradientes transversales para las máquinas de imágenes por resonancia magnética Giroimag
Year: 1998
Type: article
Abstract: The 16-arc saddle coil to yield a magnetic field gradient has been used extensively in magnetic resonance imaging machines. In recently works, this configuration had been optimized assuming only the degrees of freedom that introduces the main arc: ampere turn, axial position and angular width. In this work. assuming the main and the outer arc as a basic unity, we introduce a new degree of freedom. The coordinates of the vertices of each turn are moved in order to minimize the root mean square of the difference between the magnetic field generated by the processed configuration and the target magnetic field. We obtain a magnetic field gradient coil that is 1.92 times more uniform and produces 1.1 times less inductance. We use the Simplex method as a mathematical tool.
Source: Revista Mexicana de Física
Author: Hector Sanchez Lopez
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: diamond
OpenAlex ID: W10075132