Work

Electrical Impedance Tomography of Electrolysis

Year: 2015
Type: article
Abstract: The primary goal of this study is to explore the hypothesis that changes in pH during electrolysis can be detected with Electrical Impedance Tomography (EIT). The study has relevance to real time control of minimally invasive surgery with electrolytic ablation. To investigate the hypothesis, we compare EIT reconstructed images to optical images acquired using pH-sensitive dyes embedded in a physiological saline agar gel phantom treated with electrolysis. We further demonstrate the biological relevance of our work using a bacterial E.Coli model, grown on the phantom. The results demonstrate the ability of EIT to image pH changes in a physiological saline phantom and show that these changes correlate with cell death in the E.coli model. The results are promising, and invite further experimental explorations.
Source: PLoS ONE
Language: English
Open Access status: gold
Estimated APC paid: $2,382
OpenAlex ID: W1011963042
FWCI: 0.599
Cites: 49
Cited by: 12