Fabrication of intense neutron sources for medical applications

Year: 1975
Type: article
Abstract: Simulated sources containing $sup 252$Cf equivalents of 0.1 to 1.0 milligrams were prepared. Samarium was used as the simulant in a modified chemical plating technique similar to that used to prepare palladium-californium oxide cermet for industrial applications. The length of the platinum-10 percent iridium doubly encapsulated source with its protective sheath is 0.545 in. (14.1 mm). Outside dia of the source, including its sheath, is 0.109 in. (2.8 mm). Existing ''Brachytrons'' can accommodate this source form. This capsule system will withstand internal gas pressures from helium due to alpha decay and fission gases from a 1 mg $sup 252$Cf source after ten years if the source is subjected to a maximum temperature of 800$sup 0$C, the theoretical temperature of an accidental fire. Under these conditions the safety factor is 3. The capsule system is being tested with tracer amounts of $sup 252$Cf to ensure that it will withstand adverse service conditions as well as tests specified for Special Form Materials. (auth)
Authors: A.R. Boulogne, Valentin Walker
Language: English
Open Access status: green
OpenAlex ID: W104166257
FWCI:
Cites: 3
Cited by:
Related to: 10