Refractive indices with Haag-Streit Lenstar biometer

Year: 2014
Type: article
Abstract: Purpose: To estimate refractive indices used with the Lenstar biometer. Methods: Axial lengths of model eyes were determined using an IOLMaster biometer and a Lenstar; comparing these lengths gave an overall eye index for the Lenstar. Using the Lenstar Graphical User interface, we determined that boundaries between media could be manipulated so that there were opposite changes in optical pathlength on either side of the boundary and specified changes in distances determined the ratios of media indices. These ratios were combined with the overall eye index to estimate indices. Results: The IOLMaster and Lenstar produced axial length estimates to within ±0.01 mm. Estimations of group refractive indices were 1.340, 1.341, 1.415 and 1.354 for cornea, aqueous, lens and overall eye, respectively. The aqueous and lens indices, but not those for the cornea, are similar to schematic eye indices and reasonable lens indices. Conclusion: The Lenstar appears to use different refractive indices for different ocular media.