Year: 2015
Type: article
Abstract: Acetaminophen (paracetamol) interferes with continuous glucose monitor (CGM) sensing, resulting in falsely elevated CGM glucose values in both sensors currently approved by the U.S. Food and Drug Administration (FDA). In amperometric glucose biosensors, particularly those measuring hydrogen peroxide, acetaminophen’s phenolic moiety is oxidized at the sensing electrode, producing an electrochemical signal not related to glucose (1). Limited published data exist documenting the magnitude of the effect of acetaminophen on CGM glucose (2), especially in the outpatient setting with contemporary sensor technology. Currently, the FDA recommends that insulin dosing decisions are based on blood glucose (BG) meter values, not CGM glucose values. Given the common use of acetaminophen, its interference with CGM sensing has significant clinical implications for patients who use CGM. To better understand this effect, we performed an acetaminophen challenge as part of an outpatient study designed to investigate the potential challenges to closed-loop systems, which use CGM sensor …