Fully Integrated Analog Front-end for a 2-electrode ECG device

Year: 2011
Type: dissertation
Abstract: C ardiovascular diseases are leading cause of deaths worldwide. With increasing graying population and limited health infrastructure there is a need for portable and implantable ECG devices to diagnose these diseases early on. In general an ECG device requires three or more electrodes, which need to be applied to the body via a gel, to obtain a satisfactory reading. This puts a constraint on the portability of the device. This thesis deals with designing a fully-integrated 10 bit analog front-end i.e. an instrumentation amplifier and an ADC, specif-ically, for a two-electrode ECG device. Integrated ECG read-out circuits have to deal with two challenges mainly viz. obtaining a high common-mode rejection ratio (CMRR) and integrating large time-constants on the chip. Firstly, all sources of interference which affect an ECG reading is studied. It is shown that for a portable and integrated read-out circuit, a high CMRR is obtained from the fact that the device will be floating and hence the circuit itself need not have a high CMRR. Existing techniques for integrating large time-constants are presented and compared. It is shown