Cause and Amelioration of MRI-Induced Heating Through Medical Implant Lead Wires

Year: 2014
Type: conference-paper
Abstract: The RF fields present in magnetic resonance imaging (MRI) scanners can induce hazardous heating in patients wearing medical implants. The inherent design and locale of deep brain stimulators (DBS) and spinal cord stimulators (SCS) make them particularly susceptible. We apply antenna concepts and use electromagnetic (EM) simulation to explain the phenomenon and anticipate its sensitivity to lead wire length. We anticipate that a DC resistance of less than 50 Ω/m and an RF impedance of more than 1:23kΩ/m would be required for a safe electrode for SCS use. We investigate the possibility of manipulating wire conductivity and diameter in order to use the skin depth effect to achieve a safe electrode. The effect of the thickness and permittivity of insulation surrounding the wires is explored.