Equipamento de biotelemetria para monitoramento em tempo real do teste da caminhada de seis minutos
Year: 2004
Type: conference-abstract
Abstract: This work aims to develop a portable radio communication system of restricted radiation for digital biotelemetry of short distance. It was used in the 6-Minute Walk Test (6MWT) in patients with chronic pulmonary obstructive disease and pulmonary hypertension. The hemoglobin peripheral saturation (SpO2) and the cardiac frequency (CF) were monitored in real time. The Industrial, Scientific and Medical (ISM) band was used with a carrier frequency of 916MHz and power transmission of 0,75mW. This system was designed to operate in half duplex mode using NRZ Manchester codification and automatic repeat request error protocol with a CRC-16 code for error detection. The maximum baud rate was 115.2 kbps. The system consists of three parts: portable unit (Master), stationary unit (Slave) and software for real time visualization. The portable unit receives the parameters from the pulse oximeter which are transmitted through a radio-frequency link. The stationary unit communicates with the computer using a standard serial RS-232 port. The tests demonstrated the effectiveness of the developed biotelemetry system, without movement restriction of the user during the 6MWT. The precision of SpO2 is ±3% (±1 Standard Deviation in 0%-100% range), the CF precision is ±3% and the Frame Error Rate (FER) < 10 (0,01%).
Language: Portuguese
Topic: Wireless Body Area Networks
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1484902608