Abstract: A calibration method aimed for the differential nonlinearity (DNL) error of the Low Energy X-ray Instrument (LE) onboard the Hard X-ray Modulation Telescope (HXMT) is presented, which is independent with electronic systems used as testing platform and is only determined by the analog-to-digital converter (ADC) itself. Exploring this method, ADCs that are used within the flight model phase of HXMT-LE can be calibrated individually and independently by a non-destructive and low-cost way, greatly alleviating the complexity of the problem. As a result, the performance of the energy spectrum can be significantly improved, further more, noise reduced and resolution enhanced.
Source: arXiv (Cornell University)
Authors: Bo, Lu, Yan-Hong, Fu, Yong, Chen, Wei-Wei, Cui, Yu-Sa, Wang, Yu-Peng, Xu, Guo-He, Yin, Yan-Ji, Yang, Yi, Zhang, Jia, Huo, Wei, Li, Tian-Xiang, Cheng, Wei, Hu, Da-Wei, Han, Mao-Shun, Li, Juan, Wang, Yu, Wang, Xiao-Yang, Liu, Yue, Zhu, Zhong-Yi, Zhao, Ya, Zhang
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1523638239