Work

The Potentiometric Urea Biosensor Using Chitosan Membrane

Year: 2010
Type: article
Abstract: Potentiometric urea biosensor development is based on urea hydrolysis by urease resulted CO2. The biosensor is used chitosan membrane and the H3O+ electrode as a transducer. The research was studied of effecting pH and membrane thickness to the biosensor performance. The best biosensor performance resulted at pH = 7.3 and 0.2 mm of membrane thickness. The biosensor has a Nerntian factor 28.47 mV/decade; the concentration range is 0.1 up to 6.00 ppm; and the limit of detection is 0.073 ppm. The response time of this biosensor is 280 seconds, efficiency 32 samples and accuracy 94% up to 99%. Keywords: biosensor, potentiometry, urea, chitosan membrane
Language: English
Field: Chemistry
Open Access status: gold
Estimated APC paid: $300
OpenAlex ID: W1541796126
FWCI: 0.595
Cites: 9
Cited by: 16