Assessing the MR compatibility of dental retainer wires at 7 Tesla

Year: 2013
Type: article
Abstract: PURPOSE: To determine the MR compatibility of common dental retainer wires at 7 Tesla in terms of potential RF heating and magnetic susceptibility effects. METHODS: Electromagnetic simulations and experimental results were compared for dental retainer wires placed in tissue-mimicking phantoms. Simulations were then performed for a human model with wire in place. Finally, image quality was assessed for different scanning protocols and wires. RESULTS: Simulations and experimental data in phantoms agreed well, with the length of the wire correlating to maximum heating in phantoms being approximately 47 mm. Even in this case, no substantial heating occurs when scanning within the specific absorption rate (SAR) guidelines for the head. Image distortions from the most ferromagnetic dental wire were not significant for any brain region. CONCLUSION: Dental retainer wires appear to be MR compatible at 7 Tesla.