A working with a Silicon Position Sensitive Detector (dE%E calibration and isotopes separation)
Year: 2005
Type: preprint
Abstract: A method for the calibration of a Silicon Position Sensitive Detector is described. From the combination of the signals collected at the ends of the resistive electrode, energy and angle of the hitting particle are reconstructed with a procedure which straightforwardly accounts for different amplifier chain gains and dc offset effects.
Source: arXiv (Cornell University)
Authors: Ostashko, V., Tumino, A., Romano, S.
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1545744757