ESTIMATION OF MEASUREMENT UNCERTAINTY IN THE DETERMINATION OF Fe CONTENT IN POWDERED TONIC FOOD DRINK USING GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY

Year: 2010
Type: article
Abstract: The evaluation of uncertainty measurement in the determination of Fe content in powdered tonic food drink using graphite furnace atomic absorption spectrometry was carried out. The specification of measurand, source of uncertainty, standard uncertainty, combined uncertainty and expanded uncertainty from this measurement were evaluated and accounted. The measurement result showed that the Fe content in powdered tonic food drink sample was 569.32 µg/5g, with the expanded uncertainty measurement ± 178.20 µg/5g (coverage factor, k = 2, at confidende level 95%). The calibration curve gave the major contribution to the uncertainty of the final results. Keywords: uncertainty, powdered tonic food drink, iron (Fe), graphite furnace AAS
Language: English
Subfield: Instrumentation
Open Access status: gold
Estimated APC paid: $300
OpenAlex ID: W1553440387
FWCI:
Cites: 1
Cited by: 2