Abstract: Scanning electron micrographs at high magnification (100,000x and up) are distorted by motion of the sample during image acquisition, a phenomenon called drift. We propose a method for correcting drift distortion in images obtained on scanning electron and other scanned-beam microscopes by registering a series of images to create a drift-free composite. We develop a drift-distortion model for linear drift and use it as a basis for an affine correction between images in the sequence. The performance of our correction method is evaluated with simulated datasets and real datasets taken on both scanning electron and scanning helium-ion microscopes; we compare performance against translation only correction. In simulation, we exhibit a 12.5 dB improvement in SNR of our drift-corrected composite compared to a non-aligned composite, and a 3 dB improvement over translation correction. A more modest 0.4 dB improvement is measured on the real image sets compared to translation correction alone.