Year: 2015
Type: article
Abstract: As relative systematic frequency uncertainties in trapped-ion spectroscopy are approaching the low 10−18 range, motional frequency shifts account for a considerable fraction of the uncertainty budget. Micromotion, a driven motion fundamentally connected to the principle of the Paul trap, is a particular concern in these systems. In this article, we experimentally investigate at this level three common methods for minimizing and determining the micromotion amplitude. We develop a generalized model for a quantitative application of the photon-correlation technique, which is applicable in the commonly encountered regime where the transition linewidth is comparable to the rf drive frequency. We show that a fractional frequency uncertainty due to the 2nd-order Doppler shift below |Δν/ν|=1×10−20 can be achieved. The quantitative evaluation is verified in an interleaved measurement with the conceptually simpler resolved sideband method. If not performed deep within the Lamb-Dicke regime, a temperature-dependent offset at the level of 10−19 is observed in resolved sideband measurements due to sampling of intrinsic micromotion. By direct comparison with photon-correlation measurements, we show that the simple to implement parametric heating method is sensitive to micromotion at the level of |Δν/ν|=1×10−20 as well.
Source: Journal of Applied Physics
DOI: 10.1063/1.4930037
Authors: J. Keller, H. L. Partner, T. Burgermeister, T. E. Mehlstäubler
Institution: Physikalisch-Technische Bundesanstalt
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1584843495