Scanning laser imaging of the retina. Basic concepts and clinical applications
Year: 2011
Type: dissertation
Abstract: Fundus autofluorescence dynamics by photopigment bleaching.During recording of shortwavelength fundus autofluorescence on a dark-adapted, healthy eye a continuous increase of the autofluorescence signal can be observed.More photons pass the light-adapted retina than the dark-adapted retina.Consequently, an increasing number of fluorophores of the retinal pigment epithelium are excited during bleaching and a stronger signal is observed.Therefore, fundus autofluorescence should always be measured on the bleached retina so that reliable fluorescence signal intensities are obtained.