Compressive fatigue properties of acrylic bone cement for vertebroplasty
Year: 2013
Type: conference-paper
Abstract: INTRODUCTION: Acrylic bone cements based on poly(methyl methacrylate) (PMMA) are commonly used for the fixation of joint prostheses as well as for vertebroplasty. Fatigue properties of the cement are important to the long-term success of the implant. The ASTM F2118-03 standard– originally developed for cements used in joint fixations – specifies a fully reversed tensioncompression test. However, the same standard is commonly used for testing of cements intended for vertebroplasty, despite the fact that vertebrae mainly experience compressive loads. Until now, purely compressive fatigue studies of PMMA are scarce [1, 2]. Studies on vertebroplastic cement, which may contain a higher amount of radiopacifier, are none. The aim of this study was to evaluate the compressive fatigue properties of vertebroplastic cements, as well as the effect of frequency on these properties. Due to the time consuming character of fatigue tests – the standard cites a frequency of 2 Hz – there is a desire for increasing the frequency to higher values.
Authors: Ingrid Ajaxon, Cecilia Persson
Language: English
Subfield: Surgery
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W1594813587