Air-stable reference material for measurement of the oxygen-to-metal (O/M) ratio of nuclear fuel
Year: 1983
Type: article
Abstract: As a reference material for the oxygen-to-metal (O/M) ratio of nuclear oxide fuel, pellets prepared by calcining powder blends of titanium suboxide and thorium oxide in hydrogen atmosphere have desirable features of controllable O/M-ratio values, long-term stability in air, absence of plutonium, and high resistance to fracture. Under the conditions of the O/M-ratio measurement method of heating 800 to 900/sup 0/C in an atmosphere of 6% hydrogen - 94% helium containing water vapor, the titanium suboxide oxidizes stoichiometrically to titanium dioxide. Pellets have been prepared having O/M-ratio values equivalent to values of 1.94 to 1.98 for 3/1 uranium/plutonium mixed-oxide fuel, and have been used effectively for calibration and quality-control surveillance of an automated, predictive-thermogravimetric analyzer being developed by the Hanford Engineering Development Laboratory.
Authors: Ceri Hammond, R.E. Honnell, J.E. Rein
Institution: Los Alamos National Laboratory
Language: English
Subfield: Materials Chemistry
Field: Materials Science
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1595590201