Miniaturization of a Magnetically Levitated Axial Flow Blood Pump
Year: 2010
Type: article
Abstract: This article introduces a unique miniaturization process of a magnetically levitated axial flow blood pump from a functional prototype to a pump suitable for animal trials. Through COMSOL three-dimensional finite element analysis and experimental verification, the hybrid magnetic bearings of the pump have been miniaturized, the axial spacing between magnetic components has been reduced, and excess material in mechanical components of the pump was reduced. Experimental results show that the pump performance was virtually unchanged and the smaller size resulted in the successful acute pump implantation in calves.
Source: Artificial Organs
Institutions: Rochester Institute of Technology, Mayo Clinic
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1618916525