On the Use of the Scintillation Counter as a Gamma Ray Spectrometer, Using Stilbene Crystal
Year: 1953
Type: article
Abstract: A preliminary investigation on the use of the scintillation counter as a gamma ray spectrometer has been reported here, using the 01 game crystal stilbene and the IP28 photomultiplier tube Pulse height distribution curves have been obtained for the Co60 source with a thin lead radiator and a thin aluminum foil respectively. Co60 emits two gamma rays of energies 1.17 and 1.33 Mey. ; an attempt has been made to explain the formation of the peaks in the distribution curves in the light of the different processes through which gamma rays lose energy in the crystal material; the positions of the peaks agree well with the positions expected from theoretical considerations The results have been discussed with mentions of the reports of different workers in this line. The electronic circuits and the experimental arrangements designed for the, present purpose have been described in detail
Authors: Sagar K. Dhar, Shashwati Sen
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W164355435