Work

Image quality assessment of head CT: Control charts as an useful instrument

Year: 2013
Type: article
Abstract: Themain objective of this research was to determine if the diagnostic image, acquired by CT scan, meets the quality criteria previously established for head CT exams. A total of 360 Head Computed Tomography exams were analyzed, using a checklist. For data collection, quality criteria were created, organized into four criteria groups, consisting of multiple items that must appear in the images studies. After data processing, a large number of non-conformexaminations were identified in than more than 50% of the sample. We concluded the main causes of these results are: the “incorrect or incomplete positioning”, the "lack of name of the radiographer and “motion artefacts”. Therefore it is essential to implement a checklist for a systematic evaluationof procedures.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W16459292
FWCI:
Cites: 21
Cited by:
Related to: 10