Improved Bloch‐Siegert based B1 mapping by reducing off‐resonance shift
Year: 2013
Type: article
Abstract: An MRI method based on the Bloch-Siegert (BS) shift phenomenon was recently proposed as a fast and precise way to map a radio frequency (RF) transmit field (B1(+) field). For MRI at high field, the mapping sensitivity of this approach was limited by tissue heating associated with a BS irradiation pulse. To mitigate this, we investigated the possibility of lowering the off-resonance frequency of this pulse since theoretical analysis indicated that the sensitivity of Bloch-Siegert based B1(+) mapping could be substantially improved when irradiating closer to resonance. Using optimized irradiation pulse shape and gradient crushers to minimize direct excitation effects, in vivo experiments on human brains at 7 T confirmed improved sensitivity with this approach. Improved sensitivity translated into an 80% reduction in B1(+) estimation errors without increasing tissue heating. Published 2013. This article is a U.S. Government work and is in the public domain in the USA.
Source: NMR in Biomedicine
DOI: 10.1002/nbm.2920
Authors: Qi Duan, Peter van Gelderen, Jeff Duyn
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
Funder: National Institutes of Health
Award: G4988861893
OpenAlex ID: W1700549028