Work

Определение Т-2 токсина методом поверхностного плазмонного резонанса

Year: 2009
Type: article
Abstract: A possibility of application of optimal immune biosensor based on surface plasmon resonance for express determination of trichotecen mycotoxin T2 in solution has been investigated. Two variants of T2 toxin determination by surface plasmon resonance immunosensor method have been proposed. It was shown opportunity of «direct» determination of T2 with the detection limit of about 20 ng/ml and for «competitive» mode with the detection limit on the level of about 1 ng/ml. The effectiveness of application of such organic compounds as polyelectrolytes and thiols for sensible surface modification of SPR spectrometer for T2 analysis in solutions was studied. Direct determination by using dodecanethiol and polyelectrolyte for surface modification allows to determine T2 toxin at levels of 5 ng/ml and 2 ng/ml accordingly in case of using both poly and monoclonal antibodies.
Language: Russian
Subfield: Plant Science
Open Access status: diamond
OpenAlex ID: W194621248
FWCI:
Cites:
Cited by:
Related to: 8