Transcutaneous measurement of blood Po2(tcPo2) — Method and application in perinatal medicine

Year: 1973
Type: article
Abstract: We have previously shown that it is possible to measure Po2 transcutaneously by means of an electrode according to CLARK [1], This Po2 corre-lates with the arterial values both in newborns and in adults [2, 5, 8, 10, 11]. In order to obtain reproducible and representative tcPo2 measure-ments, attention must be paid to several factors, such äs the shape of the electrode, the size of the platinum cathodes, the thickness of the electrode membranes and the local "arterializa-tion " of the skin at the site of measurement. 1. Types of arterialization In our earlier studies, arterialization was induced by a nicotinic acid derivate*. At that time moni-toring was limited to a maximum of one hour because the tcPo2 values feil due to a reduction in the degree of vasodilatation. Constant tcPo2 readings in the newborn infant and in the adult only became possible when local vasodilatation was achieved by using heat. This heat is supplied by the electrode itself and the energy required f or keeping the tempera-ture constant is used äs a relative measurement of local blood perfusion. With this device tcPo2 may be monitored for several hours. Clinical investigations with the heated electrode have been done both in adults and in newborn infants [6, 7]. The present paper describes the Finalgon ® (Anasco GmbH, Germany). Curriculum vitae