Work

Column-type multi-component force transducers and their evaluation for dynamic measurement

Year: 2008
Type: article
Abstract: We describe the dynamic evaluation of multi-component force–moment transducers. Two of these transducers, which can measure three force components, F x , F y and F z , and three moment components, M x , M y and M z , were developed. One has a solid cylindrical sensing element and the other has a hollow cylindrical shape. They were calibrated statically with a force–moment calibration machine in the Physikalisch-Technische Bundesanstalt (PTB), Germany. We evaluated the dynamic characteristics of the force–moment transducers by using a shaker system and a multi-channel dynamic analyser system. The frequency range of the normal force component F z is very wide up to 1 kHz, the test frequency in this study. However, the frequency ranges for the transverse force components and moment components are restricted to about 200 Hz and 130 Hz respectively. The column-type force–moment transducer has a high stiffness and so its frequency range is wider than that of a binocular-type force–moment transducer.
Language: English
Open Access status: bronze
OpenAlex ID: W1968490791
FWCI: 0.749
Cites: 7
Cited by: 12
Related to: 10