High-speed imaging of human retina in vivo with swept-source optical coherence tomography
Year: 2006
Type: article
Abstract: We present the first demonstration of human retinal imaging in vivo using optical frequency domain imaging (OFDI) in the 800-nm range. With 460-muW incident power on the eye, the sensitivity is 91 dB at maximum and >85 dB over 2-mm depth range. The axial resolution is 13 mum in air. We acquired images of retina at 43,200 depth profiles per second and a continuous acquisition speed of 84 frames/s (512 A-lines per frame) could be maintained over more than 2 seconds.