Work

Double-illumination photoacoustic microscopy

Year: 2012
Type: article
Abstract: Recent developments of optical-resolution photoacoustic microscopy (OR-PAM) have improved its spatial resolution and imaging speed. However, the penetration depth of OR-PAM is still limited to ~1 mm in tissue, owing to the strong tissue scattering. Here, we have developed double-illumination PAM (DI-PAM), which illuminates the sample from both top and bottom sides simultaneously. Through phantom and in vivo experiments, we have demonstrated for thin targets that DI-PAM has a penetration depth of ~2 mm in tissue at 532 nm and a focal zone of 260 μm, both significant improvements over traditional reflection or transmission-mode OR-PAM.
Language: English
Open Access status: bronze
OpenAlex ID: W1972329969
FWCI: 4.844
Cites: 14
Cited by: 65
Related to: 10