Parallel imaging for NMR microscopy at 14.1 Tesla

Year: 2005
Type: article
Abstract: Parallel imaging techniques using arrays of mutually decoupled coils have become standard on almost all clinical imaging systems. Such techniques also have great potential for high-field magnetic resonance (MR) microscopy, where measurement times are usually long and susceptibility artifacts can be severe. However, it is technically very challenging to design efficient high-frequency phased arrays for small-diameter, vertical-bore magnets, especially since standard decoupling methods, such as impedance mismatched preamplifiers, cannot be easily integrated. A four-coil phased array was constructed for microimaging at 600 MHz, and sensitivity encoding (SENSE) and generalized autocalibrating partially parallel acquisitions (GRAPPA) reconstructions of spin-echo and echo-planar images of the mouse brain were performed to reduce imaging time and susceptibility artifacts, respectively.