Year: 1982
Type: conference-abstract
Abstract: Standing waves in the ear canal cause substantial difficulties in the assessment of high-frequency (8000–20 000 Hz) hearing thresholds because of uncertainties in the specification of the acoustic stimulus. A calibration procedure is proposed for estimating the sound prcssure p1 at the inner end of the ear canal by measuring the poles of the impulse response at the entrance to the ear canal when an acoustic source is coupled directly to the canal through a short tube. This calculation is based on the fact that the transfer function from the source to p1 is an all-pole function. As a step towards implementation of this procedure, frequencies and bandwidths of ear-canal resonances for a number of ears have been measured. The data show that these resonances have bandwidths as small as 300 Hz, and deviations from equal spacing that are usually no greater than 10%. The pressure p1 has been calculated from the appropriate all-pole transfer function, and shows systematic and predictable differences with measured peak sound pressures in the outer 1 to 1.5 cm of the ear canal, as expected on the basis of measurements with models. [Supported by a contract from NINCDS.]