Work

Fast scanning coaxial optoacoustic microscopy

Year: 2012
Type: article
Abstract: The hybrid nature of optoacoustic imaging might impose limitations on concurrent placement of optical and ultrasonic detection components, especially in high resolution microscopic applications that require dense arrangements and miniaturization of components. This hinders optimal deployment of the optical excitation and ultrasonic detection paths, leading to reduction of imaging speed and spatial resolution performance. We suggest a compact coaxial design for optoacoustic microscopy that allows optimizing both the light illumination and ultrasonic detection parameters of the imaging system. System performance is showcased in phantoms and in vivo imaging of microvasculature, achieving real time operation in two dimensions and penetration of 6 mm into optically dense human tissues.
Language: English
Open Access status: gold
Estimated APC paid: $1,913
OpenAlex ID: W1984745957
FWCI: 5.059
Cites: 18
Cited by: 69
Related to: 10