Work

Comparison between simulation results and DEXA investigation of the bone remodelling after implanting a cementless long stem hip prosthesis

Year: 2013
Type: article
Abstract: Bone remodelling around a femoral prosthesis is one of the main reasons of aseptic loosening of the implant's stem.The difference in stiffness between the bone tissue and the prosthesis leads to unload the periprosthetic bone and thus to decrease the bone mineral density (BMD), which is known as stress shielding.This paper represents a comparison between measured changes in (BMD) in a periprosthetic Femur with a cementless Bicontact stem by prospective dualenergy X-ray absorptiometry study (DEXA)and calculated changes of the (BMD) using a finite element method (FEM).As a result, it is determined that the total deviation between the two investigations is about 18%.
Language: English
Subfield: Surgery
Field: Medicine
Open Access status: bronze
OpenAlex ID: W1985244240
FWCI:
Cites: 9
Cited by:
Related to: 10