Abstract: Optical pacing has been demonstrated to be a viable alternative to electrical pacing in embryonic hearts. In this study, the feasibility of optically pacing an adult rabbit heart was explored. Hearts from adult New Zealand White rabbits (n = 9) were excised, cannulated and perfused on a modified Langendorff apparatus. Pulsed laser light (λ = 1851 nm) was directed to either the left or right atrium through a multimode optical fiber. An ECG signal from the left ventricle and a trigger pulse from the laser were recorded simultaneously to determine when capture was achieved. Successful optical pacing was demonstrated by obtaining pacing capture, stopping, then recapturing as well as by varying the pacing frequency. Stimulation thresholds measured at various pulse durations suggested that longer pulses (8 ms) had a lower energy capture threshold. To determine whether optical pacing caused damage, two hearts were perfused with 30 µM of propidium iodide and analyzed histologically. A small number of cells near the stimulation site had compromised cell membranes, which probably limited the time duration over which pacing was maintained. Here, short-term optical pacing (few minutes duration) is demonstrated in the adult rabbit heart for the first time. Future studies will be directed to optimize optical pacing parameters to decrease stimulation thresholds and may enable longer-term pacing.
Source: Biomedical Optics Express
DOI: 10.1364/boe.4.001626
Institutions: Case Western Reserve University, Cleveland Clinic
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,913
OpenAlex ID: W1988615695