Combination of 2D sensitivity encoding and 2D partial fourier techniques for improved acceleration in 3D contrast‐enhanced MR angiography
Year: 2005
Type: article
Abstract: Sensitivity encoding (SENSE) and partial Fourier (PF) techniques both reduce MRI acquisition time. Two-dimensional SENSE uses coil sensitivities to unfold aliasing in the phase/slice-encoding plane. One-dimensional PF and homodyne reconstruction are routinely applied in the frequency/phase-encoding plane to compensate for nonsampled k-space of the presumed real magnetization. Recently, a modified 3D elliptical centric acquisition was proposed to facilitate 2D-PF and homodyne reconstruction on an undersampled phase/slice-encoding plane. In this work we hypothesized that this 2D-PF technique can be combined with 2D-SENSE to achieve a greater acceleration factor than what each method can provide separately. Reconstruction of data whereby SENSE and PF are applied along the same axes is described. Contrast-enhanced MR angiography (CE-MRA) results from experiments using four receiver coils in phantom and volunteer studies are shown. In 11 volunteer studies, the SENSE-PF-homodyne technique using sevenfold acceleration (4x SENSE, 1.7x PF) consistently provided high-diagnostic-quality images with near 1-mm isotropic resolution in acquisition times of <20 s.
Source: Magnetic Resonance in Medicine
DOI: 10.1002/mrm.20742
Institution: Mayo Clinic
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W1990217036