X-ray–optical cross-correlator for gas-phase experiments at the Linac Coherent Light Source free-electron laser

Year: 2012
Type: article
Abstract: X-ray–optical pump–probe experiments at the Linac Coherent Light Source (LCLS) have so far been limited to a time resolution of 280 fs fwhm due to timing jitter between the accelerator-based free-electron laser (FEL) and optical lasers. We have implemented a single-shot cross-correlator for femtosecond x-ray and infrared pulses. A reference experiment relying only on the pulse arrival time information from the cross-correlator shows a time resolution better than 50 fs fwhm (22 fs rms) and also yields a direct measurement of the maximal x-ray pulse length. The improved time resolution enables ultrafast pump–probe experiments with x-ray pulses from LCLS and other FEL sources.
Language: English
Open Access status: hybrid
OpenAlex ID: W2000216091
FWCI: 5.901
Cites: 19
Cited by: 92
Related to: 10