Effects of Tip-Applicator Positioning on Recorded Temperature Profiles in Cryoablation Procedures Estimated by Finite Element Modeling
Year: 2013
Type: article
Abstract: Cryoablation protocols can be assessed using temperature distributions in the ablated region obtained from numerical simulations.During in-vivo studies of cardiac cryoablation the exact positioning of the used applicator is not always known, leading to possible deviations between experiments and simulations.To assess the effects of alternative tip-applicator positionings on the temperature profile of the tip-applicator finite element models of different alignments (horizontal and vertical) and insertion depths (1.5-4.5 mm for the vertical alignment) were simulated.Only small deviations in the temperature profiles were found between the simulated variants that mainly depend on the tissue contact area.Quasi-stationary temperatures were in a close range with an absolute difference below 4 • C.
Authors: Michael Handler, Roland Kienast, Gerald Fischer, Michael Seger
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2000604055