Novel Design for Centrifugal Counter-current Chromatography: Iii. Saw Tooth Column
Year: 2010
Type: article
Abstract: The toroidal coil using an equilateral triangular core and zigzag pattern column have improved both retention of the stationary phase and peak resolution of the conventional toroidal coil in centrifugal counter-current chromatography. To further improve the retention of stationary phase and peak resolution, a novel saw tooth column was designed and the performance of the system was evaluated at various flow rates. The results indicated that both retention of the stationary phase and peak resolution were improved as the flow rate was decreased and at a flow rate of 0.005 ml/min the resolution is remarkably increased. Modification of the tubing called flat-twisted tubing further improved the peak resolution without increasing the column pressure. With a decreased column length at a capacity of about 0.2 ml, resolution of the saw tooth column was 1.02.
Authors: Yi Yang, Haji Akber Aisa, Yoichiro Ito
Language: English
Subfield: Analytical Chemistry
Field: Chemistry
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W2001396352