Year: 1988
Type: article
Abstract: An international comparison in the pressure range 20 - 100 MPa has been carried out under the auspices of the high-pressure working group of the Comité Consultatif pour la Masse et les grandeurs apparentées (CCM) of the Comité International des Poids et Mesures (CIPM). The Standards Laboratories of 13 countries have participated in this comparison, which took place during the period 1981 - 1985. This paper presents a résumé of the comparison. The transfer standard used was an oil-operated pressure balance. Each laboratory determined the effective area of the piston-cylinder assembly of this balance in the pressure range 20 - 100 MPa. The results of the measurement of the effective area extrapolated to zero applied pressure agreed within 204 parts per million (ppm) for all 13 laboratories (9 laboratories agreed within 53 ppm). The results of the measurement of the effective area at 100 MPa agreed within 414 ppm for all 13 laboratories (8 laboratories agreed within 78 ppm).
Source: Metrologia
Authors: J C Legras, S L Lewis, G F Molinar
Language: English
Subfield: Statistics, Probability and Uncertainty
Field: Decision Sciences
Domain: Social Sciences
Open Access status: bronze
Funder: Chung Shan Medical University
OpenAlex ID: W2002348937