A novel fluorescence lifetime imaging system that optimizes photon efficiency
Year: 2007
Type: article
Abstract: Fluorescence lifetime imaging (FLIM) is a powerful microscopy technique for providing contrast of biological and other systems by differences in molecular species or their environments. However, the cost of equipment and the complexity of data analysis have limited the application of FLIM. We present a mathematical model and physical implementation for a low cost digital frequency domain FLIM (DFD-FLIM) system, which can provide lifetime resolution with quality comparable to time-correlated single photon counting methods. Our implementation provides data natively in the form of phasors. On the basis of the mathematical model, we present an error analysis that shows the precise parameters for maximizing the quality of lifetime acquisition, as well as data to support this conclusion. The hardware and software of the proposed DFD-FLIM method simplifies the process of data acquisition for FLIM, presents a new interface for data display and interpretation, and optimizes the accuracy of lifetime determination.
DOI: 10.1002/jemt.20540
Authors: Ryan A. Colyer, Claudia Lee, Enrico Gratton
Institution: University of California, Irvine
Language: English
Subfield: Biophysics
Domain: Life Sciences
Open Access status: green
OpenAlex ID: W2005893031