Work

An echo-shifted gradient-echo MRI method for efficient diffusion weighting

Year: 1999
Type: article
Abstract: A segmented magnetic resonance imaging (MRI) method is introduced with time-efficient diffusion weighting resulting in total imaging times similar to those of single-shot methods. The approach is based on the principles of echo shifting with a train of observations (PRESTO) MRI sequence. The time efficiency of the sequence is based on the use of diffusion gradient pulses that also serve to shift the echo train to the next TR period, resulting in TE > TR. Each diffusion gradient is therefore used twice, for dephasing one set of spins as well as rephasing a second set of spins. Diffusion weighting and acquisition are thus achieved simultaneously. The sequence is validated in vitro and in vivo on rat kidney.
Language: English
Field: Medicine
Open Access status: bronze
OpenAlex ID: W2006227704
FWCI: 0.903
Cites: 43
Cited by: 19
Related to: 10