Work

Controlled Comparison of BacT/Alert MB System, Manual Myco/F Lytic Procedure, and Isolator 10 System for Diagnosis of Mycobacterium tuberculosis Bacteremia

Year: 2011
Type: article
Abstract: We compared the performance of the BacT/Alert MB system, that of the manual Bactec Myco/F Lytic procedure, and that of the Isolator 10 lysis-centrifugation system in the detection of Mycobacterium tuberculosis bacteremia. Mean times to detection were 16.4 days for BacT/Alert MB versus 20.0 days for Myco/F Lytic, 16.5 days for BacT/Alert MB versus 23.8 days for Isolator 10, and 21.1 days for Bactec Myco/F Lytic versus 22.7 days for Isolator 10. There were no significant differences in yields. The mean (range) magnitude of mycobacteremia was 30.0 (0.4, 90.0) CFU/ml and was correlated with the time to positivity in the BacT/Alert MB system (r = -0.4920). M. tuberculosis bacteremia was detected more rapidly in a continuously monitored liquid blood culture system, but the mean time to positivity exceeded 3 weeks.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W2007073808
FWCI: 2.376
Cites: 27
Cited by: 41
Related to: 10