Multiband accelerated spin‐echo echo planar imaging with reduced peak RF power using time‐shifted RF pulses
Year: 2013
Type: article
Abstract: PURPOSE: To evaluate an alternative method for generating multibanded radiofrequency (RF) pulses for use in multiband slice-accelerated imaging with slice-GRAPPA unaliasing, substantially reducing the required peak power without bandwidth compromises. This allows much higher accelerations for spin-echo methods such as SE-fMRI and diffusion-weighted MRI where multibanded slice acceleration has been limited by available peak power. THEORY AND METHODS: Multibanded "time-shifted" RF pulses were generated by inserting temporal shifts between the applications of RF energy for individual bands, avoiding worst-case constructive interferences. Slice profiles and images in phantoms and human subjects were acquired at 3 T. RESULTS: For typical sinc pulses, time-shifted multibanded RF pulses were generated with little increase in required peak power compared to single-banded pulses. Slice profile quality was improved by allowing for higher pulse bandwidths, and image quality was improved by allowing for optimum flip angles to be achieved. CONCLUSION: A simple approach has been demonstrated that significantly alleviates the restrictions imposed on achievable slice acceleration factors in multiband spin-echo imaging due to the power requirements of multibanded RF pulses. This solution will allow for increased accelerations in diffusion-weighted MRI applications where data acquisition times are normally very long and the ability to accelerate is extremely valuable.
Source: Magnetic Resonance in Medicine
DOI: 10.1002/mrm.24719
Institution: University of Minnesota
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2007294293
