Work

Sensitive Amplified Immunoenzymometric Assays (IEMA) for Human Insulin and Intact Proinsulin

Year: 1992
Type: article
Abstract: Immunoenzymometric assays (IEMAs) for human insulin and intact proinsulin were developed using the amplification system developed by Johannsson et al. (Clin. Chim. Acta 148 (1985) 119-124) for the detection of the enzyme alkaline phosphatase. The detection limit of the assays was 0.8 pmol/l for proinsulin and 0.8 pmol/l for insulin whereas it was 1.8 pmol/l and 2.3 pmol/l respectively for the homologous immunoradiometric assays (IRMA). These assays are superior to immunoradiometric assays in terms of sensitivity, shelf-life of the labelled antibody and suitability for automation.
Language: English
Open Access status: green
OpenAlex ID: W2009055216
FWCI: 1.348
Cites: 15
Cited by: 53