Year: 2013
Type: article
Abstract: The effect of grain refinement of commercial purity titanium by equal channel angular pressing (ECAP) on its mechanical performance and bone tissue regeneration is reported. In vivo studies conducted on New Zealand white rabbits did not show an enhancement of biocompatibility of ECAP-modified titanium found earlier by in vitro testing. However, the observed combination of outstanding mechanical properties achieved by ECAP without a loss of biocompatibility suggests that this is a very promising processing route to bioimplant manufacturing. The study thus supports the expectation that commercial purity titanium modified by ECAP can be seen as an excellent candidate material for bone implants suitable for replacing conventional titanium alloy implants.
Source: BioMed Research International
DOI: 10.1155/2013/914764
Authors: Yuri Estrin, Hyoun-Ee Kim, Rimma Lapovok, Hoi Pang Ng, Ji-Hoon Jo
Institutions: Seoul National University, Monash University
Language: English
Subfield: Materials Chemistry
Field: Materials Science
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $2,680
Award: G2198712800
OpenAlex ID: W2015888867