Work

The uncertainty of the phase-correction in sphere-diameter measurements

Year: 2012
Type: article
Abstract: The calculation of the phase correction and the associated uncertainty in the measurement of sphere diameters carried out by optical interferometers is of particular interest. With this view, a Gaussian-beam tracing method was utilized for application to the sphere-interferometer at the National Metrology Institute of Japan as it was used to measure the volume of the 28 Si spheres in the determination of the Avogadro constant. Monte Carlo simulations taking account also of possible interferometer misalignments reveal a correction and its uncertainty of 0.609(139) nm.
Source: Metrologia
Language: English
Open Access status: bronze
OpenAlex ID: W2016235408
FWCI: 2.179
Cites: 13
Cited by: 21
Related to: 10