Correction of the Calibration of the 3-wavelength Particle Soot Absorption Photometer (3λ PSAP)
Year: 2010
Type: erratum
Abstract: During the Reno Aerosol Optics Study (RAOS) in June 2002, the prototype of the 3-wavelength Particle Soot Absorption Photometer (3λ PSAP) was calibrated and a correction scheme was subsequently published (Virkkula, A., Ahlquist, N. C., Covert, D. S., Arnott, W. P., Sheridan, P. J., Quinn, P. K., and Coffman, D. J. (2005). Modification, Calibration and a Field Test of an Instrument for Measuring Light Absorption by Particles. Aerosol. Sci. Technol., 39:68–83). Unfortunately, the 3 λ PSAP flow rate had not been corrected during that analysis. The factors in the transmittance correction change after reanalysis with flow-corrected data. Reanalysis of the data also resolves an apparent discrepancy between the 1 λ PSAP and the 3 λ PSAP, and the two instruments now agree well.
Source: Aerosol Science and Technology
Author: Aki Virkkula
Institution: University of Helsinki
Language: English
Subfield: Atmospheric Science
Field: Earth and Planetary Sciences
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W2021712569
