Syringe-pump-induced fluctuation in all-aqueous microfluidic system implications for flow rate accuracy
Year: 2013
Type: article
Abstract: We report a new method to display the minute fluctuations induced by syringe pumps on microfluidic flows by using a liquid-liquid system with an ultralow interfacial tension. We demonstrate that the stepper motor inside the pump is a source of fluctuations in microfluidic flows by comparing the frequencies of the ripples observed at the interface to that of the pulsation of the stepper motor. We also quantify the fluctuations induced at different flow rates, using syringes of different diameters, and using different syringe pumps with different advancing distances per step. Our work provides a way to predict the frequency of the fluctuation that the driving syringe pump induces on a microfluidic system and suggests that syringe pumps can be a source of fluctuations in microfluidic flows, thus contributing to the polydispersity of the resulting droplets.
Source: Lab on a Chip
DOI: 10.1039/c3lc51176f
Authors: Zida Li, Sze Yi Mak, Alban Sauret, Ho Cheung Shum
Institutions: University of Hong Kong, Princeton University
Language: English
Subfield: Fluid Flow and Transfer Processes
Field: Chemical Engineering
Domain: Physical Sciences
Open Access status: green