New fabrication of high-frequency (100-MHz) ultrasound PZT film kerfless linear array [Correspondence]

Year: 2013
Type: article
Abstract: The paper describes the design, fabrication, and measurements of a high-frequency ultrasound kerfless linear array prepared from hydrothermal lead zirconate titanate (PZT) thick film. The 15-μm hydrothermal PZT thick film with an area of 1 × 1 cm, obtained through a self-separation process from Ti substrate, was used to fabricate a 32-element 100-MHz kerfless linear array with photolithography. The bandwidth at -6 dB without matching layer, insertion loss around center frequency, and crosstalk between adjacent elements were measured to be 39%, -30 dB, and -15 dB, respectively.
Language: English
Open Access status: green
OpenAlex ID: W2026365929
FWCI: 1.656
Cites: 25
Cited by: 53