Work

Passive Aerosol Sampler. Part I: Principle of Operation

Year: 2001
Type: article
Abstract: A method has been developed to estimate average concentrations and size distributions with a miniature passive aerosol sampler. To use the passive sampler, one exposes it to an environment for a period of hours to weeks. The passive sampler is intended to monitor ambient, indoor, or occupational aerosols and has potential utility as a personal sampler. The sampler is inexpensive and easy to operate and is capable of taking long-term samples to investigate chronic exposures. After sampling, the passive sampler is covered and brought to the lab. Scanning electron microscopy (SEM) and automated image analysis are used to count and size collected particles with dp
Authors: Jeff Wagner, David Leith
Language: English
Open Access status: bronze
OpenAlex ID: W2027948434
FWCI: 2.832
Cites: 52
Cited by: 131
Related to: 9