Work

A method for determining the spring constant of cantilevers for atomic force microscopy

Year: 1996
Type: article
Abstract: Cantilevers fabricated by means of micromachining techniques are usually used for atomic force microscopy. In this paper, the spring constant of an atomic force microscope (AFM) cantilever is determined by using a large-scale cantilever. Since the spring constant of the large-scale cantilever is calibrated accurately, the spring constant of the AFM cantilever is determined precisely by measuring the deflections of both cantilevers simultaneously using heterodyne interferometry. The slope of the force curve gives the spring constant of the AFM cantilever. It is not necessary to measure the dimensions of the AFM cantilever in the proposed method. Although this method is simple, the spring constant of the AFM cantilever is obtained accurately.
Language: English
Open Access status: bronze
OpenAlex ID: W2028653557
FWCI: 0.883
Cites: 15
Cited by: 317
Related to: 10